- Getinge 8666 User Manual
- Getinge Service Manuals
- Getinge Washer Disinfector 46 Series
Getinge 46-series washer-disinfectors combine hygienic and ergonomic solutions with well-proven technology for maximum user safety. Choose between two, four and five levels, depending on the type and quantity of items you need to process. Smoke/fire was observed coming from a getinge model 46-4 washer disinfector during use. The washer was approx 35-40 minutes into a 70 minute cycle when staff observed smoke/fire. The user's turned off the power to the unit and sprayed with a.
-
510(k) | | | DeNovo | | | Registration & Listing | | | Adverse Events | | | Recalls | | | PMA | | | HDE | | | Classification | | | Standards |
CFR Title 21 | | | Radiation-Emitting Products | | | X-Ray Assembler | | | Medsun Reports | | | CLIA | | | TPLC |
|
GETINGE DISINFECTION AB GETINGE 46-SERIES WASHER DISINFECTOR | Back to Search Results |
| Model Number 46-4 | Device Problems Fire; Smoking | Event Date 08/13/2013 | Event Type Malfunction | Event Description | Smoke/fire was observed coming from a getinge model 46-4 washer disinfector during use. The washer was approx 35-40 minutes into a 70 minute cycle when staff observed smoke/fire. The user's turned off the power to the unit and sprayed with a dry powder extinguisher. | Manufacturer Narrative | Device was evaluated remotely (not returned to manufacturer) via investigation reports and photographs from getinge (b)(4). , a service unit located in (b)(4). It was concluded that the overheat protection system failed/delayed response to an overheat condition. A voluntary device correction of getinge model 46-4 washer disinfectors was reported to u. S. Fda on (b)(4) 2013 (recall number not yet assigned). This correction was initiated to address two conditions. A potential trigger for overheating; (root cause) - inadequate overheat protection system component. A component (shunt trip) selected for the overheat protection system does not meet the required parameters (control voltage) where/when overheating occurs. The correction entails - inspecting the setting of a software function that can be utilized during servicing of a unit to skip wash phases. This is a password protected function but if left in the enabled state, a user can inadvertently activate the heating element with no water in the chamber. The device correction action includes inspecting that this function is not enabled. The installation of a newer design of overheat protection system that utilizes a heating element with integral thermal fuses that meet the required overheat parameters. Recall number not yet assigned. | Search Alerts/Recalls |
|
|
Getinge 8666 User Manual
New Search | Submit an Adverse Event Report |
Getinge Service Manuals
Getinge Washer Disinfector 46 Series
Type of Device | WASHER DISINFECTOR |
---|
Manufacturer (Section D) | GETINGE DISINFECTION AB | ljungadalsgatan 11 | vaxjo 351 15 | SW 351 15 |
|
---|
Manufacturer (Section G) | GETINGE DISINFECTION AB | ljungadalsgatan 11 | vaxjo 351 15 | SW 351 15 |
|
---|
Manufacturer Contact | ann wheeler, us agent | 1777 e henrietta rd | rochester , NY 14623 | 5852725036 |
|
---|
MDR Report Key | 3368406 |
---|
Report Number | 9616031-2013-00004 |
---|
Device Sequence Number | 1 |
---|
Product Code | MDZ |
---|
Report Source | Manufacturer |
---|
Source Type | Company Representative |
---|
Reporter Occupation |
---|
Remedial Action | Repair |
---|
Type of Report | Initial |
---|
Report Date | 08/14/2013 |
---|
1 Device Was Involved in the Event |
---|
0 PatientS WERE Involved in the Event: |
---|
Date FDA Received | 09/13/2013 |
---|
Is This An Adverse Event Report? | No |
---|
Is This A Product Problem Report? | Yes |
---|
Device Operator |
---|
Device MODEL Number | 46-4 |
---|
Was Device Available For Evaluation? | No |
---|
Is The Reporter A Health Professional? | Yes |
---|
Event Location | Hospital |
---|
Date Manufacturer Received | 08/14/2013 |
---|
Was Device Evaluated By Manufacturer? | No |
---|
Is The Device Single Use? | No |
---|
Is this a Reprocessed and Reused Single-Use Device? | No |
---|
Type of Device Usage | Invalid Data |
---|
|
|