- Getinge 8666 User Manual
- Getinge Service Manuals
- Getinge Washer Disinfector 46 Series
Getinge 46-series washer-disinfectors combine hygienic and ergonomic solutions with well-proven technology for maximum user safety. Choose between two, four and five levels, depending on the type and quantity of items you need to process. Smoke/fire was observed coming from a getinge model 46-4 washer disinfector during use. The washer was approx 35-40 minutes into a 70 minute cycle when staff observed smoke/fire. The user's turned off the power to the unit and sprayed with a.
-
| 510(k) | | | DeNovo | | | Registration & Listing | | | Adverse Events | | | Recalls | | | PMA | | | HDE | | | Classification | | | Standards |
| CFR Title 21 | | | Radiation-Emitting Products | | | X-Ray Assembler | | | Medsun Reports | | | CLIA | | | TPLC |
|

| GETINGE DISINFECTION AB GETINGE 46-SERIES WASHER DISINFECTOR | Back to Search Results |
| | Model Number 46-4 | | Device Problems Fire; Smoking | | Event Date 08/13/2013 | | Event Type Malfunction | | Event Description | Smoke/fire was observed coming from a getinge model 46-4 washer disinfector during use. The washer was approx 35-40 minutes into a 70 minute cycle when staff observed smoke/fire. The user's turned off the power to the unit and sprayed with a dry powder extinguisher. | | Manufacturer Narrative | Device was evaluated remotely (not returned to manufacturer) via investigation reports and photographs from getinge (b)(4). , a service unit located in (b)(4). It was concluded that the overheat protection system failed/delayed response to an overheat condition. A voluntary device correction of getinge model 46-4 washer disinfectors was reported to u. S. Fda on (b)(4) 2013 (recall number not yet assigned). This correction was initiated to address two conditions. A potential trigger for overheating; (root cause) - inadequate overheat protection system component. A component (shunt trip) selected for the overheat protection system does not meet the required parameters (control voltage) where/when overheating occurs. The correction entails - inspecting the setting of a software function that can be utilized during servicing of a unit to skip wash phases. This is a password protected function but if left in the enabled state, a user can inadvertently activate the heating element with no water in the chamber. The device correction action includes inspecting that this function is not enabled. The installation of a newer design of overheat protection system that utilizes a heating element with integral thermal fuses that meet the required overheat parameters. Recall number not yet assigned. | | Search Alerts/Recalls |
|
|

Getinge 8666 User Manual
| New Search | Submit an Adverse Event Report |
Getinge Service Manuals

Getinge Washer Disinfector 46 Series
| Type of Device | WASHER DISINFECTOR |
|---|
| Manufacturer (Section D) | | GETINGE DISINFECTION AB | | ljungadalsgatan 11 | | vaxjo 351 15 | | SW 351 15 |
|
|---|
| Manufacturer (Section G) | | GETINGE DISINFECTION AB | | ljungadalsgatan 11 | | vaxjo 351 15 | | SW 351 15 |
|
|---|
| Manufacturer Contact | | ann wheeler, us agent | | 1777 e henrietta rd | | rochester , NY 14623 | | 5852725036 |
|
|---|
| MDR Report Key | 3368406 |
|---|
| Report Number | 9616031-2013-00004 |
|---|
| Device Sequence Number | 1 |
|---|
| Product Code | MDZ |
|---|
| Report Source | Manufacturer |
|---|
| Source Type | Company Representative |
|---|
| Reporter Occupation |
|---|
| Remedial Action | Repair |
|---|
| Type of Report | Initial |
|---|
| Report Date | 08/14/2013 |
|---|
| 1 Device Was Involved in the Event |
|---|
| 0 PatientS WERE Involved in the Event: |
|---|
| Date FDA Received | 09/13/2013 |
|---|
| Is This An Adverse Event Report? | No |
|---|
| Is This A Product Problem Report? | Yes |
|---|
| Device Operator |
|---|
| Device MODEL Number | 46-4 |
|---|
| Was Device Available For Evaluation? | No |
|---|
| Is The Reporter A Health Professional? | Yes |
|---|
| Event Location | Hospital |
|---|
| Date Manufacturer Received | 08/14/2013 |
|---|
| Was Device Evaluated By Manufacturer? | No |
|---|
| Is The Device Single Use? | No |
|---|
| Is this a Reprocessed and Reused Single-Use Device? | No |
|---|
| Type of Device Usage | Invalid Data |
|---|
|
|